个性化文献订阅>期刊> Sensors And Actuators A-Physical
 

Skin penetration and fracture strength testing of silicon dioxide microneedles

  作者 Khanna, P; Flam, BR; Osborn, B; Strom, JA; Bhansali, S  
  选自 期刊  Sensors And Actuators A-Physical;  卷期  2011年170-1-2;  页码  180-186  
  关联知识点  
 

[摘要]This research aims to validate the utility of silicon dioxide microneedles as a fluid delivery medium. Circular as well as square lumen microneedles were fabricated via MEMS processing. Successful penetration into skin was verified by measuring insertion force using a load cell and also by performing histology of the skin. Axial as well as shear fracture strength analysis of the microneedles was done to quantify usage limits. Fracture limits of circular and square shaped silicon dioxide needles show reduced strength of square needles: which is pronounced in the case of shear fracture. The average axial fracture limit of circular needles was about 3.15 gf, and was 2.08 gf for square needles. The average shear fracture limit of circular needles was 2.21 gf, and 0.65 gf for square ones. (C) 2011 Elsevier B.V. All rights reserved.

 
      被申请数(0)  
 

[全文传递流程]

一般上传文献全文的时限在1个工作日内