个性化文献订阅>期刊> OPTICS EXPRESS
 

Application of cavity ring-down spectroscopy to the Boltzmann constant determination

  作者 Sun, YR; Pan, H; Cheng, CF; Liu, AW; Zhang, JT; Hu, SM  
  选自 期刊  OPTICS EXPRESS;  卷期  2011年19-21;  页码  19993-20002  
  关联知识点  
 

[摘要]The Boltzmann constant can be optically determined by measuring the Doppler width of an absorption line of molecules at gas phase. We propose to apply a near infrared cavity ring-down (CRD) spectrometer for this purpose. The superior sensitivity of CRD spectroscopy and the good performance of the near-ir lasers can provide ppm (part-per-million) accuracy which will be competitive to present most accurate result obtained from the speed of sound in argon measurement. The possible influence to the uncertainty of the determined Doppler width from different causes are investigated, which includes the signal-to-noise level, laser frequency stability, detecting nonlinearity, and pressure broadening effect. The analysis shows that the CRD spectroscopy has some remarkable advantages over the direct absorption method proposed before. The design of the experimental setup is presented and the measurement of C(2)H(2) line near 0.8 mu m at room temperature has been carried out as a test of the instrument. (C) 2011 Optical Society of America

 
      被申请数(0)  
 

[全文传递流程]

一般上传文献全文的时限在1个工作日内