An electrochemical glucose biosensor exploiting a polyaniline grafted multiwalled carbon nanotube/perfluorosulfonate ionomer-silica nanocomposite

  作者 Gopalan, AI; Lee, KP; Ragupathy, D; Lee, SH; Lee, JW  
  选自 期刊  BIOMATERIALS;  卷期  2009年30-30;  页码  5999-6005  
  关联知识点  
 

[摘要]A glucose biosensor was fabricated with loading of glucose oxidase (GOx) into a new organic-inorganic hybrid nanocomposite. The preparation involves formation of silica network into a Nafion (per-fluorosulfonate ionomer) and subsequent loading of polyaniline grafted multiwalled carbon nanotubes (MWNT-g-PANI) onto Nafion-silica nanocomposite. Field emission scanning electron microscopy (FE-SEM) of Nafion-silica/MWNT-g-PANI composite reveals the presence of spherical silica particles (sizes in the range 250 nm-1 mu m) and tubular MWNT-g-PANI particles. Chronoamperometry and cyclic voltammetry were used to evaluate the performance of biosensor towards glucose. The Nafion-silica/MWCNT-g-PANI/GOx biosensor exhibited a linear response to glucose in the concentration range of 110 mm with a correlation coefficient of 0.9972, good sensitivity (5.01 mu A/mM), a low response time (similar to 6 s), repeatability (R.S.D value of 2.2%) and along-term stability. The presence of silica network within Nafion and MWNT-g-PANI synergistically contributes to the performance of the biosensor towards the electrochemical detection of glucose. (C) 2009 Elsevier Ltd. All rights reserved.

 