Design and development of a 3D scanning MEMS OCT probe using a novel SiOB package assembly - art. no. 125005

  作者 Xu, Y; Singh, J; Premachandran, CS; Khairyanto, A; Chen, KWS; Chen, N; Sheppard, CJR; Olivo, M  
  选自 期刊  Journal of Micromechanics and Microengineering;  卷期  2008年18-12;  页码  25005-25005  
  关联知识点  
 

[摘要]A MEMS optical coherence tomography (OCT) probe prototype was developed using a unique assembly based on silicon optical bench (SiOB) methodology. The probe is formed by integrating a three-dimensional (3D) scanning micromirror, gradient refractive index ( GRIN) lens and optical fiber on SiOB substrates having prefabricated self-aligned slots. The two-axis scanning micromirror is based on electrothermal actuation with required voltage less than 2 V for mechanical deflections up to 17 degrees. The optical probe was enclosed within a biocompatible, transparent and waterproof polycarbonate tube with a view of in vivo diagnostic applications. The diameter of the miniature probe is less than 4 mm and the length of its rigid part is about 25 mm. The probe engineering and proof of concept of the probe were demonstrated by obtaining en face and three-dimensional OCT images of an IR card used as a standard sample.

 
      被申请数(0)  
 

[全文传递流程]

一般上传文献全文的时限在1个工作日内