Batch and Size-Exclusion Chromatographic Characterization of Ultra-high Molar Mass Sodium Hyaluronate Containing Low Amounts of Strongly Scattering Impurities by Dual Low Angle Light Scattering/Refractometric Detection

  作者 Porsch, B; Laga, R; Konak, C  
  选自 期刊  Journal of Pharmacology and Experimental Therapeutics;  卷期  2008年31-20;  页码  3077-3093  
  关联知识点  
 

[摘要]

It is shown that very pure sodium hyaluronate (HA) contains small amounts of strongly scattering impurities not detectable on mass scale by refractometric detection, but clearly detectable using low angle light scattering detection during a static bulk light scattering experiment. Size filtration of its solutions does not remove these impurities, only reduces their amount depending on filter porosity. Complete removal of these particle impurities, independent of filter porosity, is achieved by hydrophobic adsorption on hydrophobic filter membranes. Using 0.1 M NaCl as a mobile phase, size exclusion chromatography (SEC) column removes the impurities by hydrophobic adsorption as well; molar masses obtained from both techniques thus agree when hydrophobic filters are used in bulk light scattering experiments. Diverse hydrodynamic flow retardation effects including slalom chromatography behavior are shown to substantially bias molar mass distributions obtained for ultra-high molar mass (UHM) HA, unless the flow rate in SEC analysis is reduced below 0.1 mL/min. Too high injected concentration (C-inj) is shown to introduce the onset of HA on column degradation. Correct polydispersity indices and molar mass distributions of UHM HA are obtained from SEC at a flow rate of 0.09 mL/min and optimized C-inj.

 
      被申请数(0)  
 

[全文传递流程]

一般上传文献全文的时限在1个工作日内