Evaluation of a Powered Intraosseous Device for Bone Marrow Sampling

  作者 Cohen, SC; Gore, JM  
  选自 期刊  Anticancer Research;  卷期  2008年28-6B;  页码  3843-3846  
  关联知识点  
 

[摘要]Background: Our objective was to evaluate a powered bone marrow aspiration device for use in diagnosing disease and monitoring disease course and medical therapy. Patients and Methods: The device was used in accordance with practice guidelines and directions for use. Data collection included insertion success, time to insertion and complications. Patient pain levels were rated 0-10 (10=extreme pain). Device operators rated ease of use of the device 0-10 (10=outstanding). Results: There were 55 patients from three centers. Successful insertion and aspiration were achieved in 54 out of 55 patients (98.1%). Mean insertion time was 4.9 +/- 3.0 seconds; significantly faster than that previously reported. There were no complications. The mean insertion pain score was 2.5 +/- 2.2 and the mean aspiration pain score was 3.7 +/- 2.5. The six operators rated the ease of use of the device at a mean score of 8.3 +/- 1.7. Conclusion: Findings suggest that the powered aspiration device is safe and effective for bone marrow aspirations.

 