个性化文献订阅>期刊> Journal of Micromechanics and Microengineering
 

Red blood cell sorting with a multi-bed microfabricated filter - art. no. 025009

  作者 Layton, BE; Lynch, B; Peter, T; Jamieson, B  
  选自 期刊  Journal of Micromechanics and Microengineering;  卷期  2012年22-2;  页码  25009-25009  
  关联知识点  
 

[摘要]A microfabricated fluidic chip for sorting red blood cells (RBCs) by size has been designed, fabricated and tested. The performance of the chip has been compared against a flow cytometer using samples from identical populations of cells, and statistically significant (p < 0.0005) differences in the measured cell size distributions were observed. The measurement paradigm reported here differs from previously demonstrated devices such as microfabricated Coulter counters or flow cytometers, in that the analysis is inherently parallel and is thus suitable for high throughput, point-of-care analysis. This study is empirical and semi-quantitative. However, important features of RBC trapping are characterized and indications for improved device design are described.

 
      被申请数(0)  
 

[全文传递流程]

一般上传文献全文的时限在1个工作日内